User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Sub PM2528 Part1 DATE: 03-Oct-93 AUTHOR: User Contributed REVISION: 1.1 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 69 NUMBER OF LINES: 150 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DRAW 8 78 2 78 67 10 33 45 1.002 DISP Turn POWER on. [DRAW10,10,F] 1.002 DISP Allow the UUT to warm up 1 hour, and proceed the test 1.002 DISP at an ambient temperature of 23 +/- 0.5 deg. Celsius. 1.003 DISP Shortcircuit the 0 and the V-Ohm terminal to check the 1.003 DISP zero point. If the zero point is not correct then press 1.003 DISP the OFFSET button. [DRAW30,60,F] 1.004 DISP Push the HIGH RES button. [DRAW30,70,F] 1.005 DISP Remove the shortcircuit, and connect testleads to the 1.005 DISP GUARD, 0, and V-Ohm terminals of the UUT. 1.005 DISP Interconnect GUARD and 0 with the terminator on the UUT. 1.006 HEAD {DC VOLTAGE TEST} 1.007 5700 200 200mV 0.005/ 0.01% 2W 2.001 5700 200 -200mV 0.005/ 0.01% 2W 3.001 5700 2 2V 0.005/ 0.01% 2W 4.001 5700 2 -2V 0.005/ 0.01% 2W 5.001 5700 20 20V 0.005/ 0.01% 2W 6.001 5700 20 -20V 0.005/ 0.01% 2W 7.001 5700 200 200V 0.005/ 0.01% 2W 8.001 5700 200 -200V 0.005/ 0.01% 2W 9.001 5700 2000 1000V 0.005/ 0.01% 2W 10.001 5700 2000 -1000V 0.005/ 0.01% 2W 11.001 5700 * S 11.002 HEAD {AC VOLTAGE TEST} 11.003 DISP Set the UUT to V AC. [DRAW10,70,F] 11.004 MESS The lowest frequency used with the AC V test is 100Hz 11.004 MESS because the 5700A has problems catching the 5700-FSC, 11.004 MESS when the frequency is lower than 100Hz. 11.005 MESS 11.006 5700 200 200mV 0.07/ 0.2% 100H 2W 12.001 5700 200 200mV 0.07/ 0.2% 10kH 2W 13.001 5700 200 200mV 0.07/ 0.4% 50kH 2W 14.001 5700 2 2V 0.07/ 0.2% 100H 2W 15.001 5700 2 2V 0.07/ 0.2% 10kH 2W 16.001 5700 2 2V 0.07/ 0.4% 50kH 2W 17.001 5700 20 20V 0.05/ 0.1% 100H 2W 18.001 5700 20 20V 0.05/ 0.1% 10kH 2W 19.001 5700 20 20V 0.05/ 0.3% 50kH 2W 20.001 5700 20 20V 0.05/ 1% 200kH 2W 21.001 5700 20 20V 0.05/ 5% 400kH 2W 22.001 ASK- W 22.002 DISP Disconnect testleads from the UUT and from the 5700. 22.002 DISP Remove the interconnection between GUARD and 0 of the 22.002 DISP UUT and connect the Optimation Power Amplifier to the 22.002 DISP 0 and V-Ohm terminals of the UUT (mind the polarity !). 22.003 DISP Switch power of the Optimation Power Amplifier on, and 22.003 DISP wait until the READY lamp lights up (ca. 1 min.). 22.004 MESS Press power amplifier ON. 22.005 5700 200 200V 0.05/ 0.1% 100H 2W 23.001 5700 200 200V 0.05/ 0.1% 10kH 2W 24.001 5700 200 200V 0.05/ 0.3% 50kH 2W 25.001 5700 2000 600V 0.05/ 0.1% 100H 2W 26.001 5700 2000 600V 0.05/ 0.1% 10kH B1 2W 27.001 MESS 27.002 5700 * S 27.003 DISP Disconnect the UUT from the Power Amplifier. 27.004 HEAD {DIRECT CURRENT TEST} 27.005 DISP Set the UUT to A DC. [DRAW10,90,F] 27.006 DISP Connect testleads to the GUARD, 0, and A terminals of 27.006 DISP the UUT, and connect them to the 5101B output. 27.006 DISP (use the V GUARD of the 5101B). 27.007 5700 20 20uA 0.02/ 0.1% 2W 28.001 5700 20 -20uA 0.02/ 0.1% 2W 29.001 5700 200 200uA 0.02/ 0.1% 2W 30.001 5700 200 -200uA 0.02/ 0.1% 2W 31.001 5700 2 2mA 0.02/ 0.1% 2W 32.001 5700 2 -2mA 0.02/ 0.1% 2W 33.001 5700 20 20mA 0.02/ 0.1% 2W 34.001 5700 20 -20mA 0.02/ 0.1% 2W 35.001 5700 200 200mA 0.02/ 0.1% 2W 36.001 5700 200 -200mA 0.02/ 0.1% 2W 37.001 DISP Interconnect GUARD and 0 on the UUT with a short wire. 37.002 5700 2 1.9A 0.02/ 0.1% 2W 38.001 5700 2 -1.9A 0.02/ 0.1% 2W 39.001 5700 * S 39.002 DISP Remove the interconnection between GUARD and 0 on the 39.002 DISP UUT. 39.003 HEAD {ALTERNATING CURRENT TEST} 39.004 DISP Set the UUT to A AC. [DRAW10,95,F] 39.005 5700 20 20uA 0.07/ 0.2% 50H 2W 40.001 5700 20 20uA 0.07/ 0.2% 1kH 2W 41.001 5700 20 20uA 0.07/ 0.4% 5kH 2W 42.001 5700 200 200uA 0.07/ 0.2% 50H 2W 43.001 5700 200 200uA 0.07/ 0.2% 1kH 2W 44.001 5700 200 200uA 0.07/ 0.4% 5kH 2W 45.001 5700 2 2mA 0.07/ 0.2% 50H 2W 46.001 5700 2 2mA 0.07/ 0.2% 1kH 2W 47.001 5700 2 2mA 0.07/ 0.4% 5kH 2W 48.001 DISP Interconnect GUARD and 0 on the UUT with a short wire. 48.002 5700 20 20mA 0.07/ 0.2% 50H 2W 49.001 5700 20 20mA 0.07/ 0.2% 1kH 2W 50.001 5700 20 20mA 0.07/ 0.4% 5kH 2W 51.001 5700 200 200mA 0.07/ 0.2% 50H 2W 52.001 5700 200 200mA 0.07/ 0.2% 1kH 2W 53.001 5700 200 200mA 0.07/ 0.4% 5kH 2W 54.001 DISP Press AUTO ranging off and press ranging UP six times 54.001 DISP (= highest range). 54.002 5700 2 1.9A 0.07/ 0.2% 50H 2W 55.001 5700 2 1.9A 0.07/ 0.2% 1kH 2W 56.001 5700 2 1.9A 0.07/ 0.4% 5kH 2W 57.001 5700 * S 57.002 DISP Press AUTO ranging on. 57.003 DISP Remove the connection between GUARD and the 0 terminal. 57.004 HEAD {TWO-WIRE RESISTANCE TEST} 57.005 DISP Set the UUT to Ohm 2W. [DRAW10,70,F] 57.006 DISP Disconnect testlead from the A terminal and connect it 57.006 DISP with the V-Ohm terminal. 57.006 DISP Disconnect the testleads from the 5101B and connect them 57.006 DISP to the 5700A output. 57.007 5700 200 190Z 0.005/ 0.03% 2W 58.001 5700 2 1.9kZ 0.005/ 0.03% 2W 59.001 5700 20 19kZ 0.005/ 0.03% 2W 60.001 5700 200 190kZ 0.005/ 0.03% 2W 61.001 5700 2 1.9MZ 0.005/ 0.1% 2W 62.001 5700 20 19MZ 0.005/ 0.1% 2W 63.001 5700 200 100MZ 0.05/ 0.2% 2W 64.001 HEAD {FOUR-WIRE RESISTANCE TEST} 64.002 DISP Set the UUT to Ohm 4W. [DRAW10,80,F] 64.003 DISP Disconnect all testleads, and connect the 4-wire 64.003 DISP testcable with probe to the UUT and to the 5700A output 64.003 DISP and sense. 64.004 DISP Interconnect GUARD and 0 with the terminator on the UUT. 64.005 5700 200 190Z 0.005/ 0.03% 4W 65.001 5700 2 1.9kZ 0.005/ 0.03% 4W 66.001 5700 20 19kZ 0.005/ 0.03% 4W 67.001 5700 200 190kZ 0.005/ 0.03% 4W 68.001 5700 2 1.9MZ 0.005/ 0.1% 4W 69.001 DISP { } 69.001 DISP { } 69.001 DISP { } 69.001 DISP { } 69.001 DISP { } 69.001 DISP { } 69.001 DISP { } 69.001 DISP { } 69.002 ASK+ W 69.003 END